Upcoming Events
Log In
Pricing
Free Trial

X-ray Collimation

HIDE
PrevNext

0:00

Now, what is X-ray collimation? The way the X-ray collimation is

0:05

achieved as follows. There are two sets of collimator

0:09

blades generally present.

0:11

One is called a round iris. It's like a camera

0:15

eye closes, open. It confirms the X-ray beam to the

0:19

image or intensifier field of view.

0:22

You say nine inches, it'll adjust it, and in addition,

0:25

collimation also have rectangular lead plates, which can be

0:29

brought in to create more of a rectangular field.

0:33

So this is the function of a collimation.

0:36

Why is this important? Because for in order to

0:40

reduce or keep the radiation dose only focused on the area of

0:43

interest and reduce patient dose unnecessarily exposed on the

0:47

healthy tissue, collimators should always be

0:51

used with live fluoro and recording mode.

0:54

I'm just showing here an example of a button which indicates which if you

0:58

move in which direction, it'll either open up the X-ray field

1:03

or closes it.

1:05

One of the advantage of the proper collimation is it reduces

1:09

scatter radiation. It reduces image quality.

1:12

In this particular right side is a two image done on the pain clinic,

1:16

where the first one is a pain clinic that's trying to

1:20

do a location of the vertebral bodies, and this is the

1:24

spine of the patient, and the field is completely open, and

1:28

that should not be the way. You can see the fuzziness

1:32

because of the large field. If we collimate to the area of

1:36

interest, only to the area, immediately the image quality will increase,

1:40

and the contrast is also increased.

1:43

So in generally, the collimation reduces scatter

1:46

radiation, improves image quality, and that's one of the reason

1:51

why collimation is very essential, and learning about it is very

1:55

essential during the practice.

1:58

This is like my colleague usually call it buttonology, and

2:02

we advise the user, such as the radiologist or the cardiologist,

2:06

to become familiar with some of the buttons on the fluoroscopy

2:10

system, especially in the interventional, where this is

2:14

almost formed like a iPad dimension kept

2:18

connected to the X-ray table. So you can see here, this is

2:22

a three-mode fluoroscopy system, which can be set at three different

2:26

dose level. That's why this is small, medium, and large.

2:29

And then you have also image intensifier button showing plus and

2:33

minus basically can go into magnify or not.

2:37

These two buttons basically are adjusted, which will

2:41

show you which direction the collimation blades are working,

2:45

as you can see on both side, or you can see together collimate to

2:49

the area of interest.

2:51

Where do we need to do a collimation?

2:54

Good example is

2:56

by doing a proper collimation, the image quality improves.

3:01

In fact, it increases the contrast

3:04

and reduces image blooming.

3:06

As shown here of a mobile C-arm image of a

3:10

lumbar spine, it is because it is open

3:13

completely, you have lot of averaging done, therefore, the image

3:17

quality has poorer image contrast.

3:20

On the other hand, by collimating to exactly to the region of interest

3:25

and remove all this extraneous need for converting X-rays into

3:29

light, the image becomes more clarified

3:33

and higher image quality compared to the situation where

3:37

the collimation is completely open.

3:40

This is a busy slide basically toward showing various physics

3:44

parameters and the comparison between radiography and

3:48

fluoroscopy. I want to draw this patient's skin

3:51

dose. This is 10 to 60 milligray per minute.

3:55

This is a fluoroscopy. Whereas in radiography, when you take one X

3:58

image, it range from 0.2 to 10 milligray.

4:03

And that's why the radiation protection principle to minimize any

4:07

risk is very essential. This particular one also shows

4:11

the tube current has wide range. For pulse, it can go even

4:14

higher. And here is the X-ray focal spot size

4:18

and the duration.

Report

Faculty

Mahadevappa Mahesh, PhD, FACR, MS, FAAPM, FACMP, FSCCT, FIOMP

Professor of Radiology and Cardiology

Johns Hopkins University School of Medicine

Tags

Physics and Basic Science

Nuclear Medicine

Fluoroscopy