Interactive Transcript
0:00
Now, what is X-ray collimation? The way the X-ray collimation is
0:05
achieved as follows. There are two sets of collimator
0:09
blades generally present.
0:11
One is called a round iris. It's like a camera
0:15
eye closes, open. It confirms the X-ray beam to the
0:19
image or intensifier field of view.
0:22
You say nine inches, it'll adjust it, and in addition,
0:25
collimation also have rectangular lead plates, which can be
0:29
brought in to create more of a rectangular field.
0:33
So this is the function of a collimation.
0:36
Why is this important? Because for in order to
0:40
reduce or keep the radiation dose only focused on the area of
0:43
interest and reduce patient dose unnecessarily exposed on the
0:47
healthy tissue, collimators should always be
0:51
used with live fluoro and recording mode.
0:54
I'm just showing here an example of a button which indicates which if you
0:58
move in which direction, it'll either open up the X-ray field
1:03
or closes it.
1:05
One of the advantage of the proper collimation is it reduces
1:09
scatter radiation. It reduces image quality.
1:12
In this particular right side is a two image done on the pain clinic,
1:16
where the first one is a pain clinic that's trying to
1:20
do a location of the vertebral bodies, and this is the
1:24
spine of the patient, and the field is completely open, and
1:28
that should not be the way. You can see the fuzziness
1:32
because of the large field. If we collimate to the area of
1:36
interest, only to the area, immediately the image quality will increase,
1:40
and the contrast is also increased.
1:43
So in generally, the collimation reduces scatter
1:46
radiation, improves image quality, and that's one of the reason
1:51
why collimation is very essential, and learning about it is very
1:55
essential during the practice.
1:58
This is like my colleague usually call it buttonology, and
2:02
we advise the user, such as the radiologist or the cardiologist,
2:06
to become familiar with some of the buttons on the fluoroscopy
2:10
system, especially in the interventional, where this is
2:14
almost formed like a iPad dimension kept
2:18
connected to the X-ray table. So you can see here, this is
2:22
a three-mode fluoroscopy system, which can be set at three different
2:26
dose level. That's why this is small, medium, and large.
2:29
And then you have also image intensifier button showing plus and
2:33
minus basically can go into magnify or not.
2:37
These two buttons basically are adjusted, which will
2:41
show you which direction the collimation blades are working,
2:45
as you can see on both side, or you can see together collimate to
2:49
the area of interest.
2:51
Where do we need to do a collimation?
2:54
Good example is
2:56
by doing a proper collimation, the image quality improves.
3:01
In fact, it increases the contrast
3:04
and reduces image blooming.
3:06
As shown here of a mobile C-arm image of a
3:10
lumbar spine, it is because it is open
3:13
completely, you have lot of averaging done, therefore, the image
3:17
quality has poorer image contrast.
3:20
On the other hand, by collimating to exactly to the region of interest
3:25
and remove all this extraneous need for converting X-rays into
3:29
light, the image becomes more clarified
3:33
and higher image quality compared to the situation where
3:37
the collimation is completely open.
3:40
This is a busy slide basically toward showing various physics
3:44
parameters and the comparison between radiography and
3:48
fluoroscopy. I want to draw this patient's skin
3:51
dose. This is 10 to 60 milligray per minute.
3:55
This is a fluoroscopy. Whereas in radiography, when you take one X
3:58
image, it range from 0.2 to 10 milligray.
4:03
And that's why the radiation protection principle to minimize any
4:07
risk is very essential. This particular one also shows
4:11
the tube current has wide range. For pulse, it can go even
4:14
higher. And here is the X-ray focal spot size
4:18
and the duration.