Upcoming Events
Log In
Pricing
Free Trial

Magnification

HIDE
PrevNext

0:00

The next part is the magnification.

0:04

Let's define the field of view. The field of view is defined

0:08

as the size of the image seen at the image receptor, in

0:12

this case, an image intensifier.

0:14

Here's an X-ray source. This is the actual field of view in

0:18

the patient, but on the image it looks magnified.

0:23

So

0:24

the size of the input phosphor of the I.I.

0:27

will determine the field of view.

0:29

The most common is a nine-inch I.I.

0:32

Also available is the larger 12 to 16-inch I.I.,

0:36

and also smaller 14-inch and 6-inches.

0:39

The actual image diameter in patient is smaller

0:43

than due to geometric magnification.

0:47

This is a large, non-magnified field of view.

0:51

This is a small magnified field of view with six inches in

0:54

diameter.

0:56

You can see so many details in this one.

0:59

Here is basically showing example of I.I.

1:02

image magnifiers, which will reduce minification.

1:06

Less minification.

1:08

Same output phosphor, but the contribution from

1:12

normal tissue was this much.

1:15

Now that is shrinking and this is shrinking.

1:19

There is what is called electronic magnification.

1:22

It reduces the field of view, it reduces the minification,

1:26

and the displayed image is magnified,

1:29

and that's electronic magnification.

1:31

But in general, for a normal operation, here is an

1:35

image field of view, which is the nine-inch I.I.

1:37

size,

1:39

and its output is usually about two to four inches.

1:43

If you reduce the field of view to small,

1:47

but the amount of X-rays required to create the image quality is the

1:51

same.

1:52

The one of the way system does it by increasing the

1:56

radiation dose to the patient, so more X-rays reach this limited

2:00

area.

2:01

That's magnification penalty with respect to AI.

2:05

Whereas in electronic magnification, as in flat panel, it can

2:09

magnify the display image

2:12

and the kind of electronic adjustment to update the

2:15

image brightness.

2:18

This particular slide showing the difference in the magnification mode,

2:22

normal versus magnified. You can see the image magnified

2:26

with much better viewing for the physician.

2:30

The beam automatically collimated to the much smaller imaging area.

2:35

Automatic brightness control will increase the radiation dose.

2:39

Another aspect is the resolution in magnification mode.

2:43

If we magnify the resolution,

2:47

here is a 23 centimeter unmagnified of a tool used

2:51

for spatial resolution.

2:53

Here is 15 centimeter, you can begin to see the object, but here is

2:56

the magnitude. At 11 centimeter, the spatial resolution

3:00

much better, but the radiation dose penalty is quite high.

3:05

So how does it compare it here? For comparison purposes,

3:09

if you require 100 X-ray units, arbitrary units to

3:13

create an image with using a nine-inch I.I.,

3:16

if we magnify to six

3:19

inches,

3:20

the radiation dose will increase by 2.25 times.

3:24

And if the magnification is go to second level,

3:28

the radiation dose will go up almost by nine times.

3:33

This is what I meant is use of smaller field of view in flat panel

3:37

detector so you can automatically magnify to a larger

3:41

point.

3:43

The intense radiation dose flat panel versus image

3:47

intensifier is shown here. With image intensifier,

3:50

usually it goes off as the field of view changes.

3:54

It goes dramatic decrease, whereas in flat panel, it's

3:58

much slower decrease.

4:00

That is captured in the following slide called the dose rate.

4:04

The I.I. fluoro dose rate increases as one over

4:08

field of view squared, which means the field of view will make a lot

4:12

of difference in the radiation dose rate.

4:15

In the flat panel dose rate, it increases only by

4:19

approximately one over the field of view.

4:22

All factors considered, flat panel dose rates are much

4:26

lower

4:27

because it has better efficiency, increased lower depth, a smaller

4:31

field of view, and the equipment configuration.

4:35

Again, this is a busy slide showing different

4:38

fluoro more available on this particular system.

Report

Faculty

Mahadevappa Mahesh, PhD, FACR, MS, FAAPM, FACMP, FSCCT, FIOMP

Professor of Radiology and Cardiology

Johns Hopkins University School of Medicine

Tags

Physics and Basic Science

Nuclear Medicine

Fluoroscopy