Interactive Transcript
0:01
Modern fluoroscopy equipment systems have various
0:05
type of configuration.
0:08
They can be
0:09
under X-ray tube table, under table X-ray tube, which is
0:13
usually in a typical radiology department, which is
0:16
radiography or fluoroscopy under the table X-ray tube.
0:21
Or the fluoroscopy system can be over the X-ray tube.
0:25
Then you have the fixed C-arm positioner.
0:28
It can be mono or a biplane system. Then you have the
0:32
mobile C-arm and, of course, the mini C-arm
0:35
also. So here is the capturing of
0:39
four different systems
0:41
showing the configuration of the equipment systems.
0:45
The most common fluoroscopy system in radiology is general
0:49
under table system.
0:51
Whereas if you move into interventional area,
0:55
such as in radiology or in cardiology, you're going to see this type of a
0:59
beastly C-arm
1:01
planted to the ground, but the C-arm shape can be rotated very easily.
1:05
Then there are system with a biplane system.
1:08
But these days, most ubiquitous fluoroscopy system in the hospital
1:13
is the mobile C-arm, which can be moved around and more
1:16
generally used in the surgical suite where it can be brought in and out
1:20
when they're doing the surgery.
1:22
So let's examine some of the typical X-ray equipment configuration.
1:28
This is the radiography fluoroscopy system, which has an under the table
1:32
X-ray tube.
1:33
When I say these type of suites or systems has both
1:37
radiography and fluoroscopy, because you can see in this room there is
1:41
a X-ray system independently. So when you
1:45
don't have to use the fluoroscopy, the room is used for taking just
1:49
X-ray or plain X-rays.
1:51
When it is used as a fluoroscopy mode, the application are
1:54
typically for GI, for GU, and
1:58
myelogram, arthrography, and even interventional.
2:03
The features are as follows. You have the X-radiography
2:07
with overhead X-ray tube, as you can show here.
2:10
Then you have a table which can be tilted based on the
2:14
operator's desire. Then you have an X-ray tube
2:17
underneath the table,
2:19
quite shielded with this particular table.
2:23
You have the image receptor here, which is the intensifier here,
2:27
which allow the user to adjust the collimation,
2:31
or you can do a fluoro recording or a
2:36
spot film device which has a photo spot and digital photo spot.
2:40
And this is the workhorse of most radiology department
2:44
in the world for a lot of the simple fluoroscopy procedure.
2:49
When it gets complicated, when you want to do longer and longer, that's when
2:53
we move into the larger side C-arm.
2:57
Another type of configuration is over the table X-ray tube.
3:01
This system you are seeing less and less these days.
3:05
Here the configuration here is the X-ray table, X-ray tube, but
3:09
underneath is the image receptor.
3:12
Let's say it has the image intensifier.
3:15
This is typically was used in urology, where the urologist used to
3:19
sit at the end of the table and position the X-ray tube to hit the
3:22
pelvic area for examining that area and the receptor.
3:27
So the radiography with the same X-ray tube, but it has an angling
3:31
X-ray tube and can be remotely controlled.
3:35
But however, there's one of the drawback of this particular system was there is a
3:38
lot of increased personal dose for the physician working in the system.
3:43
These days, we see less and less of this configuration.
3:46
But this is a biplane position. This you
3:50
typically do not see in the radiology department, but you are
3:54
seeing in interventional radiology or interventional
3:57
cardiology departments. And it is typically used for
4:01
radiography and angiography procedure,
4:04
and both in cardiology, neurointerventional, or
4:08
regular radiology interventional.
4:10
The feature is it has a C-arm sealed to the
4:14
floor
4:16
on which is there is a C-arm system here, or
4:20
it has a mounted arm where this whole
4:24
carriage is carrying another system.
4:27
So this is one system with an X-ray tube, image receptor,
4:30
X-ray tube, and the image receptor, and they can
4:34
be simultaneously operated to give a biplane
4:38
view for the user.
4:41
The most commonly used in the hospital these days,
4:45
outside of the radiology and in the radiology, is the mobile C-arm.
4:50
To give you a scale at Johns Hopkins, we have more than 60
4:53
mobile C-arm in the hospital, predominantly outside
4:57
radiology, used for lot of routine care around the
5:01
patient by orthopedics, urologist,
5:05
and everywhere, and then mostly in the surgical suite
5:08
also. So as you can see here, again, it's a compact system
5:13
which can be wheeled to the bed of the table, and that's why
5:17
the shape C is to insert it very closely into the patient
5:20
bed. There is a self-contained unit, which is a unit also with
5:24
the monitors, so you can see the X-rays
5:28
fluoroscopy images when it's operating.
5:32
These days, some of the image receptors are replaced by a flat panel
5:35
detector.
5:37
The principle behind is all the same.
5:41
One of the other type of equipment configuration of fluoroscopy is called
5:45
mini C-arm, which operates at a microampere rate.
5:49
These mini C-arms can go from the size here is about
5:53
two inches or four inches, all the way up to six-inch image
5:57
receptor.
5:58
And they are predominantly used by hand surgeons or
6:02
for pediatric cases for hand surgery or extremity surgery and
6:06
everything, and that's why you commonly do not see
6:10
in the radiology department, but more so in the operating area
6:14
by the orthopedics, hand surgeons, and so forth.
6:18
The application is mostly used for extremity.
6:21
The extremity's thickness is small, so therefore you don't need too much
6:25
X-ray output, and the exposure is also very small.
6:29
But all the radiation protection requirement also apply
6:33
for the mini C-arm, too.
6:36
These days, most of the interventional rooms are coming with a
6:40
biplane system or a single plane system,
6:43
and also with a CT scanner in place in the room.
6:47
Here is photograph of one such IR suite at Hopkins
6:51
here, which has a fluoroscopy and a
6:54
CT scanner in the suite so that the
6:57
fluoroscopist who are doing an interventional procedure can move the
7:01
table to the CT area and get a CT without having to move the
7:05
patient.
7:07
Then you might be interested in how this C-arm is inside an MR suite.
7:12
And because of that, there are way which can be brought
7:16
inside an MR suite. However, it has to be
7:20
tethered to the wall, and it cannot move beyond this
7:23
particular area, as you can see here.
7:25
Otherwise, it can be sucked into the gantry.
7:28
But it is also used outside, like in a place.