Interactive Transcript
0:00
The topic I'm going to talk today is physics of fluoroscopy.
0:05
My name is Mahesh. I'm a professor of radiology and radiological science
0:09
at the Johns Hopkins University School of Medicine, and I'm a
0:12
board-certified medical physicist.
0:15
I'm going to talk on the following outline.
0:18
In this particular part, fluoroscopy versus
0:20
radiography, equipment configuration,
0:24
X-ray generator and collimation,
0:26
fluoroscopy imaging modes,
0:29
image intensifier versus flat panel detectors,
0:33
magnification and field of view, and pulse fluoroscopy.
0:37
So let's look in the difference between fluoroscopy and
0:41
radiography.
0:43
Fluoroscopy radiation exposure rate is much lower than
0:47
radiography. That's the first thing,
0:50
and fluoroscopy is a dynamic imaging versus radiography is
0:54
a static imaging.
0:56
For example, here, for an abdominal entrance radiation
0:59
exposure rate for a fluoroscopy, it's about 45
1:03
milligray per minute.
1:06
For radiography, it's 3 milligray,
1:10
and at 200 milliseconds, that accounts to about
1:14
900 milligray a minute. So there is an order of 20 times
1:17
magnitude between radiography and fluoroscopy.
1:21
The low exposure rates is required to avoid
1:25
radiation injury to patient. We're going to discuss later why fluoroscopy
1:29
is done at such a lower radiation exposure rate.
1:33
But we don't do one frame, but we do multiple frame.
1:37
Because of that, the fluoroscopy detectors, the image
1:41
receptor, requires a very high gain.
1:44
When I say high gain, means the conversion of imaging from
1:48
X-ray exposure to actual digital image.
1:52
That receptor does all these things.
1:54
It requires a high optimal or high gain.
1:58
Compared with radiography, fluoroscopy imaging have
2:02
significantly greater noise.
2:05
It also reduced spatial resolution.
2:08
If you just compare one frame of fluoroscopy image
2:12
versus a radiographic image, the image of fluoroscopy is
2:16
slightly higher noise. However, that is integrated with the number of
2:19
frames we do.
2:21
With respect to fluoroscopy, the fluoroscopy at a typical image
2:25
receptor is, for fluoroscopy, is one line pair
2:29
per millimeter,
2:31
whereas radiography can go as high as up to 10 line
2:35
pair per millimeter.
2:37
So when I say the line pair per millimeter, that is the way the
2:41
physics we quantify the spatial
2:44
resolution in any of the X-ray imaging.
2:48
This is a schematic diagram of a fluoroscopy imaging chain.
2:53
You look here on the left-hand side is a photograph of a
2:56
mobile C-arm,
2:59
which is an image intensifier based on the top and
3:03
image flat panel detector at the bottom.
3:06
The fundamental components of an imaging chain are as
3:10
follows. One is the X-ray tube is common.
3:14
Then there is a collimator housing, which allows
3:18
user to adjust the beam size.
3:22
You have then the table where the patient is lied, and the
3:26
table is made up of a material such that it has least
3:30
amount of absorption of the X-rays.
3:33
Then radiation passing through the patient is now then captured
3:36
by an image receptor. In this case, I'm
3:40
showing an image intensifier.
3:43
At the entrance surface, there is a grid.
3:45
The purpose of the X-ray grid is to block off any
3:49
scatter radiation. Then you have the electrical, such
3:53
as optical coupling and video camera for image intensifier,
3:57
slightly different for flat panel, but these are the fundamental
4:01
components of a fluoroscopy imaging chain.
4:06
Historically, fluoroscopy began to be used right from the time when
4:10
X-rays were discovered.
4:12
Radiologists were using this even in the absence of image
4:15
intensifier, wherein the radiologist used to go into the room
4:19
of a fluoroscopy system,
4:21
turn the lights off, get their eyes adjust to the darkness, and then
4:25
move very close to the patient. And this is a
4:29
screen through which the X-ray would pass through the patient
4:32
and allow the radiologist to see a very fuzzy image.
4:37
And we can only see that the person is not wearing an apron, of course, and
4:41
then the amount of radiation required is very high.
4:43
Therefore, the patient dose was also high, and so was the exposure to
4:47
the radiologist in the early days, 1920s and '30s.
4:52
From the time an image intensifier discovered, then at the beginning, the
4:56
radiologist used to stand very close to the image intensifier because there's
5:00
only one eyepiece they used to see the image.
5:04
So early generation fluoroscopy radiologist was
5:07
required to observe directly onto the screen,
5:11
and it is only possible one person view at a time.
5:16
Later development led into a TV, monitors which multiple
5:20
people can see. These images, which are photograph images,
5:23
you can see in the historical museum in the US, at Smithsonian
5:27
Museum. These are very early stage.