Upcoming Events
Log In
Pricing
Free Trial

Fluoroscopy vs Radiography

HIDE
PrevNext

0:00

The topic I'm going to talk today is physics of fluoroscopy.

0:05

My name is Mahesh. I'm a professor of radiology and radiological science

0:09

at the Johns Hopkins University School of Medicine, and I'm a

0:12

board-certified medical physicist.

0:15

I'm going to talk on the following outline.

0:18

In this particular part, fluoroscopy versus

0:20

radiography, equipment configuration,

0:24

X-ray generator and collimation,

0:26

fluoroscopy imaging modes,

0:29

image intensifier versus flat panel detectors,

0:33

magnification and field of view, and pulse fluoroscopy.

0:37

So let's look in the difference between fluoroscopy and

0:41

radiography.

0:43

Fluoroscopy radiation exposure rate is much lower than

0:47

radiography. That's the first thing,

0:50

and fluoroscopy is a dynamic imaging versus radiography is

0:54

a static imaging.

0:56

For example, here, for an abdominal entrance radiation

0:59

exposure rate for a fluoroscopy, it's about 45

1:03

milligray per minute.

1:06

For radiography, it's 3 milligray,

1:10

and at 200 milliseconds, that accounts to about

1:14

900 milligray a minute. So there is an order of 20 times

1:17

magnitude between radiography and fluoroscopy.

1:21

The low exposure rates is required to avoid

1:25

radiation injury to patient. We're going to discuss later why fluoroscopy

1:29

is done at such a lower radiation exposure rate.

1:33

But we don't do one frame, but we do multiple frame.

1:37

Because of that, the fluoroscopy detectors, the image

1:41

receptor, requires a very high gain.

1:44

When I say high gain, means the conversion of imaging from

1:48

X-ray exposure to actual digital image.

1:52

That receptor does all these things.

1:54

It requires a high optimal or high gain.

1:58

Compared with radiography, fluoroscopy imaging have

2:02

significantly greater noise.

2:05

It also reduced spatial resolution.

2:08

If you just compare one frame of fluoroscopy image

2:12

versus a radiographic image, the image of fluoroscopy is

2:16

slightly higher noise. However, that is integrated with the number of

2:19

frames we do.

2:21

With respect to fluoroscopy, the fluoroscopy at a typical image

2:25

receptor is, for fluoroscopy, is one line pair

2:29

per millimeter,

2:31

whereas radiography can go as high as up to 10 line

2:35

pair per millimeter.

2:37

So when I say the line pair per millimeter, that is the way the

2:41

physics we quantify the spatial

2:44

resolution in any of the X-ray imaging.

2:48

This is a schematic diagram of a fluoroscopy imaging chain.

2:53

You look here on the left-hand side is a photograph of a

2:56

mobile C-arm,

2:59

which is an image intensifier based on the top and

3:03

image flat panel detector at the bottom.

3:06

The fundamental components of an imaging chain are as

3:10

follows. One is the X-ray tube is common.

3:14

Then there is a collimator housing, which allows

3:18

user to adjust the beam size.

3:22

You have then the table where the patient is lied, and the

3:26

table is made up of a material such that it has least

3:30

amount of absorption of the X-rays.

3:33

Then radiation passing through the patient is now then captured

3:36

by an image receptor. In this case, I'm

3:40

showing an image intensifier.

3:43

At the entrance surface, there is a grid.

3:45

The purpose of the X-ray grid is to block off any

3:49

scatter radiation. Then you have the electrical, such

3:53

as optical coupling and video camera for image intensifier,

3:57

slightly different for flat panel, but these are the fundamental

4:01

components of a fluoroscopy imaging chain.

4:06

Historically, fluoroscopy began to be used right from the time when

4:10

X-rays were discovered.

4:12

Radiologists were using this even in the absence of image

4:15

intensifier, wherein the radiologist used to go into the room

4:19

of a fluoroscopy system,

4:21

turn the lights off, get their eyes adjust to the darkness, and then

4:25

move very close to the patient. And this is a

4:29

screen through which the X-ray would pass through the patient

4:32

and allow the radiologist to see a very fuzzy image.

4:37

And we can only see that the person is not wearing an apron, of course, and

4:41

then the amount of radiation required is very high.

4:43

Therefore, the patient dose was also high, and so was the exposure to

4:47

the radiologist in the early days, 1920s and '30s.

4:52

From the time an image intensifier discovered, then at the beginning, the

4:56

radiologist used to stand very close to the image intensifier because there's

5:00

only one eyepiece they used to see the image.

5:04

So early generation fluoroscopy radiologist was

5:07

required to observe directly onto the screen,

5:11

and it is only possible one person view at a time.

5:16

Later development led into a TV, monitors which multiple

5:20

people can see. These images, which are photograph images,

5:23

you can see in the historical museum in the US, at Smithsonian

5:27

Museum. These are very early stage.

Report

Faculty

Mahadevappa Mahesh, PhD, FACR, MS, FAAPM, FACMP, FSCCT, FIOMP

Professor of Radiology and Cardiology

Johns Hopkins University School of Medicine

Tags

Physics and Basic Science

Nuclear Medicine

Fluoroscopy