Interactive Transcript
0:01
So having understood some of the fluoroscopic imaging
0:04
configuration, the physics principle behind these system
0:08
are pretty much the same common. So we're going to examine what is called
0:12
a fluoroscopy X-ray generator.
0:16
So fluoroscopy X-ray generator, what we try to understand is
0:20
how does this X-ray generator works,
0:23
especially when it is operating on a fluoroscopy mode
0:27
that has both a continuous fluoro
0:31
mode and also pulsed fluoro mode. We're going to discuss it more
0:35
detail in the pulsed fluoroscopy later.
0:38
When you say continuous fluoro mode means moment
0:42
you put the foot on the pedal, that's how the fluoroscopy is operated, where
0:46
there is a foot pedal where the operator such as the
0:50
radiologist or cardiologist will put pressure on the foot to
0:54
keep the X-ray on.
0:56
In the continuous fluoroscopy mode, there is a steady low
1:00
mA exposure continuously, which is
1:03
automatically able to see the image continuously.
1:07
Most of the fluoroscopy system these days are pulsed
1:10
fluoroscopy. When you say pulsed fluoro, they can
1:14
operate in different pulses, 3 to 10 millisecond
1:18
pulses at 30 pulses per second. And this
1:22
is done in order to reduce the radiation dose.
1:25
And I'm going to discuss about fluoroscopy more later.
1:28
One of the major advantage of doing pulsed fluoroscopy is it
1:31
improves the temporal resolution and also reduces the
1:35
radiation dose to the patient.
1:38
Another feature of X-ray generator is the
1:41
capability of the fluoroscopy system to have what is called
1:45
as automatic brightness control, ABC,
1:49
or automatic exposure control, also called as
1:52
AEC.
1:54
Which basically the function of this one is to keep the image
1:58
brightness constant
2:00
as system is panned over body parts of different
2:04
thickness and attenuation. You may have observed in a
2:07
fluoroscopy suite, the images changes as
2:11
brightness change. When you move a fluoroscopy system from a
2:15
thick portion area to the thinner portion, it automatically
2:19
adjusts the brightness so that it is not delivering the same
2:23
radiation dose on all area of the body
2:27
depending on the thickness. So the way the physics works behind
2:30
is, it adjusts the tube voltage and the tube current
2:34
setting as needed. The automatic brightness
2:37
control, which keeps the image brightness constant as the
2:41
system is panned over body parts of different thickness and attenuation,
2:45
is a true advantages of this particular imaging
2:49
mode. The way it does this, it varies from system to
2:53
system, but generally, it adjusts the tube current
2:57
first up to certain level, and then it adjusts the
3:01
kV to maintain the same brightness.
3:04
This is a typical diagram of an X-ray tube,
3:08
which is very useful for fluoroscopy application.
3:11
Typically, for radiography or fluoroscopy application, these
3:15
X-ray tube have bifocal focus. Means
3:20
basically it has two types of X-ray filament, which is
3:24
called bifocal system, which when it's turned on,
3:28
generates the electrons, which are then when interact with the X-ray
3:32
tube, the anode create the X-rays.
3:36
For typically for fluoroscopy, there is a small
3:40
focal spot, which is a .3 to .6 millimeter in
3:44
size, and there is also large focal spot,
3:48
1 to 1.2 millimeter for recording.
3:51
So as we discussed earlier, the focal spot makes the
3:55
determination on what's the impacts the spatial resolution.
4:00
It also impacts on heating the X-ray tube and the
4:04
life of the X-ray tube, everything.
4:06
So for fluoroscopy, small focus spot of .3 to
4:10
.6 millimeter is quite sufficient.
4:13
For a small and large focus spot when it's 1 to 1.2 is typically
4:17
used for fluoroscopy and also for recording modes.
4:22
For radiography, angiography, or interventional application,
4:26
you need very large heat capacity because you are continuously
4:30
operating the X-ray machine. And even though it's a low tube current,
4:34
but you are continuously operating, there's lot of heat generated.
4:37
Therefore, you want the system to have a X-ray tube which has a
4:41
large heat capacity,
4:43
and that's only achieved by high-speed anode rotation
4:47
and also by circulating water or oil heat exchanger with
4:51
the fan inside the X-ray tube to take the excess heat out.
4:56
There is also grid control exposure pulsing for
5:00
cine and pulse fluoroscopy, which can limit heat
5:03
capacity for smallest focal spot.
5:07
This is a schematic diagram of a collimator housing,
5:11
which sits next to the X-ray tube out of the X-ray portal
5:16
of the X-ray tube outlet where the X-rays are coming out, the
5:19
collimator housing is packed in.
5:22
And you can see here, this particular collimator housing has a
5:26
variety of material to harden the X-ray beam,
5:30
remove soft X-ray, which basically contribute to the
5:34
increased patient skin dose. And that's done, and
5:38
typically it's done either aluminum material or tungsten material or copper
5:42
material, and these are inside this one, which can be
5:45
inserted depending on the choice.