Upcoming Events
Log In
Pricing
Free Trial

Fluoroscopy X-ray Generator

HIDE
PrevNext

0:01

So having understood some of the fluoroscopic imaging

0:04

configuration, the physics principle behind these system

0:08

are pretty much the same common. So we're going to examine what is called

0:12

a fluoroscopy X-ray generator.

0:16

So fluoroscopy X-ray generator, what we try to understand is

0:20

how does this X-ray generator works,

0:23

especially when it is operating on a fluoroscopy mode

0:27

that has both a continuous fluoro

0:31

mode and also pulsed fluoro mode. We're going to discuss it more

0:35

detail in the pulsed fluoroscopy later.

0:38

When you say continuous fluoro mode means moment

0:42

you put the foot on the pedal, that's how the fluoroscopy is operated, where

0:46

there is a foot pedal where the operator such as the

0:50

radiologist or cardiologist will put pressure on the foot to

0:54

keep the X-ray on.

0:56

In the continuous fluoroscopy mode, there is a steady low

1:00

mA exposure continuously, which is

1:03

automatically able to see the image continuously.

1:07

Most of the fluoroscopy system these days are pulsed

1:10

fluoroscopy. When you say pulsed fluoro, they can

1:14

operate in different pulses, 3 to 10 millisecond

1:18

pulses at 30 pulses per second. And this

1:22

is done in order to reduce the radiation dose.

1:25

And I'm going to discuss about fluoroscopy more later.

1:28

One of the major advantage of doing pulsed fluoroscopy is it

1:31

improves the temporal resolution and also reduces the

1:35

radiation dose to the patient.

1:38

Another feature of X-ray generator is the

1:41

capability of the fluoroscopy system to have what is called

1:45

as automatic brightness control, ABC,

1:49

or automatic exposure control, also called as

1:52

AEC.

1:54

Which basically the function of this one is to keep the image

1:58

brightness constant

2:00

as system is panned over body parts of different

2:04

thickness and attenuation. You may have observed in a

2:07

fluoroscopy suite, the images changes as

2:11

brightness change. When you move a fluoroscopy system from a

2:15

thick portion area to the thinner portion, it automatically

2:19

adjusts the brightness so that it is not delivering the same

2:23

radiation dose on all area of the body

2:27

depending on the thickness. So the way the physics works behind

2:30

is, it adjusts the tube voltage and the tube current

2:34

setting as needed. The automatic brightness

2:37

control, which keeps the image brightness constant as the

2:41

system is panned over body parts of different thickness and attenuation,

2:45

is a true advantages of this particular imaging

2:49

mode. The way it does this, it varies from system to

2:53

system, but generally, it adjusts the tube current

2:57

first up to certain level, and then it adjusts the

3:01

kV to maintain the same brightness.

3:04

This is a typical diagram of an X-ray tube,

3:08

which is very useful for fluoroscopy application.

3:11

Typically, for radiography or fluoroscopy application, these

3:15

X-ray tube have bifocal focus. Means

3:20

basically it has two types of X-ray filament, which is

3:24

called bifocal system, which when it's turned on,

3:28

generates the electrons, which are then when interact with the X-ray

3:32

tube, the anode create the X-rays.

3:36

For typically for fluoroscopy, there is a small

3:40

focal spot, which is a .3 to .6 millimeter in

3:44

size, and there is also large focal spot,

3:48

1 to 1.2 millimeter for recording.

3:51

So as we discussed earlier, the focal spot makes the

3:55

determination on what's the impacts the spatial resolution.

4:00

It also impacts on heating the X-ray tube and the

4:04

life of the X-ray tube, everything.

4:06

So for fluoroscopy, small focus spot of .3 to

4:10

.6 millimeter is quite sufficient.

4:13

For a small and large focus spot when it's 1 to 1.2 is typically

4:17

used for fluoroscopy and also for recording modes.

4:22

For radiography, angiography, or interventional application,

4:26

you need very large heat capacity because you are continuously

4:30

operating the X-ray machine. And even though it's a low tube current,

4:34

but you are continuously operating, there's lot of heat generated.

4:37

Therefore, you want the system to have a X-ray tube which has a

4:41

large heat capacity,

4:43

and that's only achieved by high-speed anode rotation

4:47

and also by circulating water or oil heat exchanger with

4:51

the fan inside the X-ray tube to take the excess heat out.

4:56

There is also grid control exposure pulsing for

5:00

cine and pulse fluoroscopy, which can limit heat

5:03

capacity for smallest focal spot.

5:07

This is a schematic diagram of a collimator housing,

5:11

which sits next to the X-ray tube out of the X-ray portal

5:16

of the X-ray tube outlet where the X-rays are coming out, the

5:19

collimator housing is packed in.

5:22

And you can see here, this particular collimator housing has a

5:26

variety of material to harden the X-ray beam,

5:30

remove soft X-ray, which basically contribute to the

5:34

increased patient skin dose. And that's done, and

5:38

typically it's done either aluminum material or tungsten material or copper

5:42

material, and these are inside this one, which can be

5:45

inserted depending on the choice.

Report

Faculty

Mahadevappa Mahesh, PhD, FACR, MS, FAAPM, FACMP, FSCCT, FIOMP

Professor of Radiology and Cardiology

Johns Hopkins University School of Medicine

Tags

Physics and Basic Science

Nuclear Medicine

Fluoroscopy