Interactive Transcript
0:00
Now, let me go to the second primary factor, which is
0:04
called the tube voltage.
0:07
The tube voltage is also called as kilo volt, kV.
0:12
Because the potential difference between the anode and the
0:16
cathode, you need some energy to attract the electron to hit at
0:20
certain speed on the anode to create the X-rays.
0:24
I'm going to cover this aspect of fundamental X-ray production in a different
0:27
lecture on fundamental on radiation, radiography, and fluoroscopy.
0:32
We'll examine that over there. So right now, when we say
0:36
tube voltage in CT is all kilo voltage, kV.
0:40
What is tube voltage? This is the potential difference between the
0:44
anode and cathode of an X-ray tube.
0:47
So there is an energy to attract the electron to hit at a certain speed.
0:52
Faster the electron hit the anode,
0:55
the X-rays produced will be at a higher energy.
0:59
Therefore, the tube voltage can be looked as
1:03
to defining the quality of X-rays.
1:08
So the tube voltage will define the quality of X-ray.
1:11
The tube current will define the quantity of X-ray.
1:16
The most commonly used tube voltage in CT
1:20
is 120 kV.
1:22
Why so? Because the energy of the X-ray produced
1:26
by 120 kV is quite sufficient to pass
1:30
through the average size patient or even the thick portion of the
1:34
body and so forth. So historically,
1:38
if we look back even 15 years ago,
1:40
every CTs were done at 120 kV.
1:44
But now,
1:45
because of our understanding and way to optimize the
1:49
doses and image quality, we are now doing in different kV.
1:53
We do at a higher kV for a obese patient, lower kV for
1:58
a thin patient, and we are going down in the kV to
2:01
further optimize the radiation dose.
2:04
Especially these days,
2:07
thin and pediatric patients are done at 100
2:10
kV or 80 kV or lesser. That is the
2:14
landmark. So these kVs are calibrated on
2:18
the CT scanner by the vendor. So there is a way they do the
2:21
calibration, so that the scanner, when it is operating, provides the right
2:25
X-ray energies and so forth.
2:28
Why don't we go beyond 140 kV?
2:31
What happen is like, if the X-rays are done at a higher kV,
2:36
the X-ray energy is so great, they pass through the
2:40
body equally without providing any signal
2:43
differences. Therefore, the image quality will becomes
2:47
almost blurred. There's no difference in the contrast.
2:51
On the other hand, why don't we go very low, below 70, 60 kV?
2:56
If you go too low in a kV, the energy of the
2:59
X-rays become very soft. They just don't get penetrate
3:03
through the body. They just get absorbed on the surface of the body.
3:07
Therefore,
3:08
in diagnostic radiology, we are limited to
3:12
the range of kV we can use.
3:15
Anywhere from 50 or 60 kV to 140 to 150
3:19
kV. That is called the diagnostic energy range,
3:23
and that goes back to the principle of radiography and fluoroscopy
3:26
also. So in CT, the most common kV is
3:30
120 kV, and these are the other kVs calibrate on the
3:34
vendors.
3:35
This is a cathode, which has a filament, heats up.
3:38
That heating up is done by mA, the tube current, and the electron
3:42
produced are then attracted towards the anode by the potential
3:46
difference between these two, and that create the X-rays coming out.
3:50
As you all know, the current X-ray tubes which we use
3:55
is one of the very inefficient X-ray tube because
3:59
99% of heat is generated to produce 1% of
4:03
X-rays. And there are a lot of new technology being developed to improve
4:07
this efficiency, but we are still, most of the X-ray tubes are the same
4:11
principle.
4:12
What does the tube voltage has to do with the patient dose and image quality?
4:16
The tube voltage controls the image contrast
4:20
and also the amount of penetration.
4:23
Means the energy of the X-ray beam will
4:27
determine how much is absorbed, how much is transmitted,
4:30
allowing to develop an image contrast.
4:34
The patient dose increases with the tube voltage.
4:38
However, it is not straightforward like a tube current,
4:42
not linear. It is more like kV square.
4:46
I'm going to go into detail later. So
4:49
decreasing the tube voltage will reduce dose when all the other factors
4:53
constant. So that is where the advantage why we are going down in the tube
4:57
voltage.
4:59
So here is a scenario of the same two simulated images abdomen.
5:03
Now, I'm keeping the tube current same,
5:07
but the tube voltage is changed. Everything else is
5:11
the same. The images are displayed.
5:14
As you can see here, this is a grainier image.
5:17
This is a much more nicer image.
5:20
This was acquired at 135 kV,
5:23
which resulted in a dose of 37 milligray.
5:27
And if you put a region of interest on this area, you get what is called
5:31
as the standard deviation of the CT number, which is an
5:35
indicator of the image noise. That's eight.
5:38
Whereas here, this is done at 120 kV.
5:42
The CT volume is 29 milligray,
5:45
but there is a slight increase in the image noise. The standard deviation is 10.
5:50
One other principle in CT is the way image noise
5:54
is measured.
5:56
There are very sophisticated way to measure the image noise and
6:00
which we use it routinely for all of our research purpose.
6:03
But for simplicity case, if you take a region of
6:07
uniform region, like in the liver, and measure the CT
6:11
number of this ROI,
6:13
it will measure what is called the mean CT number in this particular region of
6:17
interest you drawAnd the standard deviation will
6:21
tell you the quantity of the image noise.
6:24
So the image noise is kind of indicated by the standard
6:29
deviation of a particular region of interest.
6:32
The standard deviation will tell whether it's high or low, whether it's
6:36
noisy or not. The derivation of these two images is like
6:41
a 12% change in tube voltage. There is only 12% change in
6:45
tube voltage,
6:46
led to almost 27% increase in
6:50
dose.
6:52
At the same time, led to 25% decrease in noise
6:56
from 120 to 135.
6:58
So then now question is, do we need
7:02
this amount of decrease in noise or you can save off this
7:06
dose? That is where the trade-off comes into picture.
7:10
So I'm showing here is for a head CT versus a tube body
7:14
CT, simulating values based on a phantom measurement.
7:18
This is a technique used, and here is showing is like for 120
7:22
kV.
7:23
If the CTDI was 50, if I go to
7:27
100, the CTDI value will be 32. That's
7:31
almost decrease of 36% dose reduction in head CT.
7:35
Head CTs are smaller size.
7:38
If you go down to 80 or 70, that number goes on even
7:41
further.
7:43
If I go to 140, that increase the dose by 44%.
7:47
So here in the body CT or a body protocol,
7:51
if it is at 120 kV, the CTDI volume
7:54
is 24 for this particular technique.
7:58
Then by going down in kV, you are saving a dose of
8:01
30%. If you go further down to 80 kV, you save a
8:05
percentage of almost 70% dose.
8:08
And that's where the current status is now with lot of the CT
8:12
protocol optimization to take into account
8:15
how we can optimize the radiation dose to the patient by adjusting
8:19
these parameters. The relationship between the tube
8:23
voltage and the patient dose is very little bit more complex.
8:28
It is tube voltage significantly reduces tube dose
8:31
by kV square.
8:33
It is not kV, but kV square.
8:36
So what I meant to say is, like a small change in the
8:40
tube voltage will have a larger impact in reducing
8:44
or increasing or reducing the radiation dose.
8:47
So you can see it, 140, 120, or 100,
8:51
and so forth. See the amount of radiation dose.
8:53
This decreased by 0.6 millisievert or 0.5
8:57
millisievert here with the 80 kV.
9:00
The most common CT application even today uses 120
9:04
kV.
9:05
However, for thin patient and infants,
9:09
we are now recommending or using 100 kV,
9:13
80 kV, and 70 kV can be used.
9:16
The advantage of using a lower tube voltage, especially
9:20
for CTs done with contrast, because it increases the
9:24
image contrast and also decreases the radiation dose.
9:28
So in a way, one of the few times when we
9:32
find dual advantage is going down to lower
9:36
tube voltage in those protocol with
9:39
contrast will have not only decreased the radiation dose to the
9:43
patient, but you are also going to increase the image contrast.
9:46
This is one of the few times there is hardly any trade-off in going
9:50
in the lower tube voltage.
9:52
At the same time, if the patient is obese or thicker anatomy, it
9:56
is advised to go into higher tube voltage to maintain the image
10:00
quality.