Interactive Transcript
0:00
Let's look a couple of trade-off very important in understanding some of these
0:04
factor.
0:05
So the effect of slice collimation on
0:09
spatial resolution and image quality.
0:12
I just want to let you know, this is assuming all the factors are the same.
0:16
Only one factor is changed, that's the slice thickness in the Z direction.
0:20
The same abdominal image here, acquired and reconstructed
0:24
0.625 and double the size, double the slice
0:28
thickness, and double the slice thickness.
0:30
So if you look in here, gradually, the primary
0:34
dose to the patient is the same. Now you are reconstructing to all
0:38
four different panel. Adjusting CT acquisition
0:42
parameter to reading image collimation is one way to help
0:46
target protocol and reduce dose. So you acquired a thinner slice,
0:50
and whatever you can, then you can reconstruct into thick slice to image
0:54
improve the contrast resolution.
0:56
So you have a very high spatial resolution, but the image quality can
1:00
be improved by simply adding the slice thicknesses.
1:05
This is another trade-off. This is the effect of tube
1:08
voltage on dose, image quality, and
1:12
contrast. So here is the study which we did in the
1:15
past, acquired at 100 kV tube voltage,
1:19
and here is the axial image and the coronal image.
1:23
And you can see here, this is done at 120
1:27
kV and coronal with contrast. See the difference between 120
1:31
and 100.
1:32
Between these two, the radiation dose is less with 100 kV,
1:36
but the coronal contrast resolution can be jeopardized.
1:40
Therefore,
1:41
to irregularly enhancing mass in the rectum is more
1:45
conspicuous due to lower tube voltage here in this area.
1:49
So at that time, you may have to increase the tube voltage to trade off.
1:53
This is with respect to iterative reconstruction.
1:56
Iterative reconstruction was done back in early days into
2:00
2006 onwards to keep the radiation dose
2:04
as low as possible. However, to mathematically do some
2:08
iterative iteration on the image reconstruction to improve the
2:12
image noise and reduce the image noise.
2:15
If you do too much mathematical averaging, the image can look
2:18
really, really like a phantom image and a plastic image.
2:22
So there is a trade-off between. Here is an image with a very
2:25
noisy image acquired at a certain
2:28
technique that on applied
2:31
iterative reconstruction was slightly about 30.
2:34
There are different vendors use different nomenclature.
2:38
This is one vendor utilizing what is called as adaptive statistical
2:42
iterative reconstruction. Using that model of the
2:46
strength of 30%, it slightly improves the image noise.
2:50
But if you go to higher and higher, you can beginning to see more like
2:54
a plastic image. So what is the effect of iterative
2:57
reconstruction? Image noise decreases with higher
3:01
percentage of iterative reconstruction used, but there's also the
3:04
effect of smoothing artifact.
3:07
So now I want to show you a series of trade-off.
3:10
One is with respect to tube current.
3:13
If the tube current is high,
3:16
you have less image noise,
3:18
better contrast resolution, and higher patient dose.
3:22
On the other hand, if the tube current is low, you have higher image
3:26
noise, lower patient dose, and decreased contrast
3:30
resolution.
3:31
As a clinician, one needs to pick where you want to go at a high tube
3:35
current or a low tube current, understanding this
3:39
in fact, which is a trade-off.
3:41
The trade-off with respect to tube voltage is as follows.
3:44
If the tube voltage is high, you have greater penetration.
3:48
It can also result in higher dose, and this is
3:52
usually desirable in obese patient to improve the signal-to-noise
3:56
ratio. Because if you don't increase the tube work voltage, the
4:00
scanner can only go up in tube current up to one point beyond
4:04
which it cannot exceed, therefore, image noise will
4:08
become predominantly very significant.
4:11
At that instance, for obese patient, it advised to increase the tube
4:14
voltage to 140 instead of 120.
4:18
If you go down on the low tube voltage, the contrast in the
4:22
image increases. There is a less patient dose,
4:26
and especially this is a desired point to select
4:30
for tube voltage for thin and pediatric subjects.
4:34
The trade-off with respect to pitch.
4:37
Pitch was defined earlier as the ratio of the table
4:41
travel to the total beam width. If the pitch is
4:45
higher, greater than one, imagine a Slinky, pull the
4:49
Slinky out, that's a pitch of greater than one, which means there are
4:53
some gap between the object, and that will result in a lower
4:57
patient dose. This will also allow faster scan time.
5:01
However, if you stretch the Slinky too much, the error in the
5:05
reconstruction can appear that can impact the image resolution and image
5:09
quality. So there is a range of pitch you can increase,
5:13
but anything greater than one pitch results in a lower
5:16
patient dose and faster scan time.
5:20
A pitch less than one is higher patient dose and
5:24
also less spiral artifact, and that's typically done
5:28
with cardiac CT we're going to study later, where if
5:32
you acquired a very low dose, very low pitch means you are
5:36
overlapping the same anatomy that can increase the patient dose.
5:41
The next trade-off is the slice thickness.
5:44
In the slice thickness, there are two ways.
5:47
If the slice thickness is large, means like five millimeter, 10 millimeter slice,
5:52
it is less noisier because there are more photon are
5:54
contributing. However, it leads to a poor spatial
5:58
resolution. It can improve this contrast resolution, higher
6:02
contrast resolution, but the spatial resolution is jeopardized.
6:06
That's where the trade-off. If on the other end, the slice thickness is very
6:10
thin, small, it has higher spatial resolution in
6:14
the Z direction and provide lower contrast
6:17
resolution, but also results in a greater noise.
6:21
So there is a trade-off between these two.