Interactive Transcript
0:00
Here are the main category of the techniques used
0:04
for obtaining dual energy CT,
0:07
and here is an example. One of the ways it's done is like a
0:10
single X-ray source with fast kV switching.
0:14
We call it as rapid switching dual energy CT technique.
0:19
The second one is a dual source with dual detector
0:23
array. We also call this a dual source dual
0:27
energy CT.
0:28
The third one is a single X-ray source with a filter,
0:32
which we call it as a twin beam dual energy CT.
0:36
Fourth one is a single source with a dual detector
0:40
layer, called as dual layer dual energy CT.
0:44
And the fifth type is basically with a regular CT scan, but
0:48
doing twice by each time setting the kV differently,
0:52
thereby you're obtaining two data set of different kV.
0:56
So here is the one which is obtained by one vendor manufacturer
1:00
called GE Healthcare. They do the dual energy CT
1:04
as follows.
1:06
It's a single source with fast kV switching.
1:09
As shown here in this particular diagram, at every
1:13
point of acquisition, there is two X-ray beams are
1:17
produced, one with 80 kV
1:20
beam and the other one is 140 kV.
1:24
Fast switching between 80 and 140 kV generates
1:28
two energy X-ray spectrum.
1:31
While doing this one, the vendor adjusts the tube
1:34
current so that for the 80 kV, the
1:38
exposure time is slightly more compared to the 140 kV
1:42
at about 35%.
1:45
The second type is called the dual source with dual detector array,
1:50
and this is quite commonly seen across the board in the clinic
1:54
because the Siemens have started doing this dual source
1:57
CT technology back in 2005 or
2:01
2006. So they have two X-ray tube with the two
2:05
detector arrays. You acquire two different image sets
2:08
simultaneously. Therefore, two different X-ray tubes allow
2:12
beam filtration and tube current adjustment to optimize the image quality.
2:17
The only thing which you need to understand is at every point, there is an
2:21
offset of one-quarter time rotation difference between the two
2:25
spectra.
2:27
Here is an example of the dual source CT.
2:30
You have two X-ray tubes installed in the CT gantry.
2:34
Simultaneously, there is two detector arrays at the opposite end.
2:38
The second detector array is slightly smaller than the first
2:42
one, and that's what this is a circle telling.
2:45
If you want to utilize the dual energy CT
2:49
protocol, the object or the patient has to be in this red
2:53
area, which covers for both the detector size of this
2:57
particular dual energy CT.
3:00
The way it's obtained here is you're obtaining two spectrum.
3:03
This is the 80 kV spectrum and 140 kV spectrum.
3:07
You can see that by having this filter, you're
3:11
separating these two separately.
3:14
The single source dual energy technique, again, is another
3:18
way the vendor is doing is they have only one
3:22
X-ray tube. However, as it's acquiring the image, there is a
3:27
tin filter passed halfway through the beam.
3:30
That way, the tin filter will distribute the energy
3:33
into two different spectrum, thereby obtaining dual
3:37
spectrum.
3:39
Another methodology obtaining dual sources or dual
3:43
energy is obtaining with the same X-ray tube,
3:47
but the detector has two layers. The top
3:50
layer will absorb the lower energy, and the bottom layer will absorb
3:54
the high energy. Thereby, we call this as a sandwich detector
3:58
type, and this modified detector array will
4:02
result in two X-ray spectrum utilized for dual energy.
4:07
The single source sequential scan is done one over the other, and
4:11
again, it can create a low or high energy spectrum.
4:15
This is less done compared to the other way.