Interactive Transcript
0:00
So I want to walk you through some of the primary factors
0:04
and how each of the factors impact both the image
0:08
quality and the radiation dose.
0:11
Among them,
0:13
one of the most fundamental primary factor is tube current.
0:17
The unit of tube current is mA, milliampere.
0:21
Ampere was a French
0:23
physicist who did a lot of fundamental work on tube current.
0:26
That's why the unit is mA.
0:30
mA is called milliampere, one-thousandth of an ampere.
0:34
Typical mAs we used in CT, if you look in
0:37
any of the patient images, you can see the mA value
0:41
displayed on which each image is. It's usually 10
0:45
mA, 100 mA, 200 mA, and so forth. That's the range we
0:49
are talking about.
0:50
So what is so important about tube current?
0:53
So if
0:55
you imagine the X-ray tube, it's like a light bulb.
0:59
When the X-ray tube is turned on, there is a filament which heats up.
1:03
When it heated up, the electrons are released.
1:07
It's called thermionic emission.
1:09
These electrons coming out from the cathode end, the filament of the X-ray
1:13
tube, now move towards the anode,
1:17
where there is a high melting point material is used, so this
1:21
electron will bombard on the anode, creating an
1:24
X-ray, such as Bremsstrahlung X-ray or characteristic X-ray.
1:29
Therefore, the tube current is in a way
1:32
directly influences the amount of X-ray produced in an X-ray tube.
1:38
Along with the tube current, we also commonly use what is called a tube
1:41
current-time product, mAs,
1:44
and that's purely multiplying the tube current with the scan time
1:48
per rotation. When I say scan time, this is the
1:52
time for the X-ray tube to go around the patient in one
1:56
rotation. It's not the total study time or the total
2:00
CT scanner time, it's just per rotation.
2:03
mAs is the tube current multiplied by the scan time per
2:07
rotation.
2:09
The factor which influence the image quality is the image
2:13
noise,
2:15
and they go in opposite direction.
2:17
So the image noises decreases with increase in tube current.
2:23
And the fundamental relationship is this one.
2:26
The increase in tube current increases radiation
2:30
dose linearly,
2:32
which means higher the tube current, higher the patient
2:36
dose.
2:37
And that principle works very straightforward with tube current.
2:42
For example, in this one, in this panel of images, you
2:46
see how the tube current can impact the image noise or image
2:49
quality.
2:51
There are four panels of images here, the same images of the abdominal
2:55
images
2:56
you have acquired at 200 mAs. That is the mA time,
3:00
the second 200 mAs. This is done at 200 mAs.
3:04
You can see the object in the kidney here.
3:07
You can also see some low contrast objects in the liver.
3:11
However, if you decrease the tube mAs to 150,
3:16
you begin to see, then notice the image becoming slightly grainier.
3:20
If you go down further to 100 or 50 mAs,
3:24
it becomes quite grainy. The image noise increases because there are
3:28
less amount of X-rays participating in image generation,
3:32
therefore, the image is noisy. And you can still see this
3:36
object,
3:37
even though it is kind of like a diffusing, is merging with the background.
3:41
But the object you were seeing in the liver is gone.
3:44
It is almost you are not able to see.
3:46
Therefore, in this panel, the renal cyst is
3:50
observed at all four radiation levels,
3:54
whereas the small vessels observed in these liver
3:57
images is gone.
4:00
This begs to the question is, which mAs should I use?
4:04
The answer is the mAs should be determined based on what you
4:08
are looking in the image.
4:11
If you're looking only for the large object here, conspicuous
4:15
object, you don't need a 200 mAs. You can use what is called
4:19
as the low mAs, therefore, resulting in lower radiation dose to the
4:23
patient. If you're looking at a very fine object, very
4:26
low subtle contrast object, then you may have to increase the tube
4:30
current higher.
4:32
I'll show you more example this when I do the cardiac CT, where
4:36
in one instance, we use very low techniques such as calcium scoring,
4:40
and we use very high technique for CT angiography.
4:44
And for the principle here is this is what the image noise can
4:47
obscure low contrast object.
4:51
So to summarize this tube current concept,
4:55
the tube current indicates the quantity of
4:59
X-rays produced.
5:00
The take home message of this one is tube current is an indicator of
5:04
the quantity of X-rays.
5:07
It is the amount of X-ray produced in X-ray tube, and the
5:10
relationship of patient dose varies linearly with tube
5:14
current. This is one of the fundamental principle behind the tube current.
5:19
On the other hand, if you decrease the tube current
5:23
or change the tube current by 50%,
5:27
that will decreases the radiation dose by 50%, but
5:31
will increase the image noise by square root of two.
5:35
This principle applies for the conventional way we
5:39
image reconstruct, where it depends on the Poisson statistics
5:43
and the number of photons contributing to the image.
5:46
So for simplicity purpose,
5:48
increase the tube current, image looks very nice.
5:51
The radiation dose is also high to the patient.
5:54
If you decrease the tube current, the radiation dose to the patient is low.
5:58
However, image becomes very grainier, and it depends on the decision
6:02
factor which of these images is sufficient for diagnosis
6:06
is the way to go towards optimization of this CT
6:09
protocol.So here are shown three different panels of
6:13
abdomen image. These are simulated images.
6:16
As you can see on the CT images, look at all the information available
6:20
here. This information is available on the scanner
6:24
depending on the way the PACS is set up.
6:26
In this particular image, it tells what kV was used on this
6:30
image, what mA was used, and what was
6:34
the scan time. This is a 0.5-second rotation time.
6:38
The number of detectors used is 32 detector times
6:42
one, one millimeter detector, and a table speed of 10 millimeter per
6:46
rotation table speed. And you also have in here
6:50
the window width and window level, which we talked
6:53
earlier when I talked about the fundamental display, the
6:57
impact of Hounsfield unit on the display.
7:00
Window level is 50, means it's slightly higher than the water, and then the
7:04
window covers all the way 150 to
7:07
-150, all the images, all the objects in this window is displayed
7:11
here.
7:13
Now, in these three panels,
7:16
this is done at three different mA,
7:19
48 mA, 69 mA, and 184.
7:24
In order to understand the variation of the tube current, the way
7:28
we physicists do is we keep all the other factors same.
7:31
So you have everything else is same, we only change the tube current.
7:35
By doing so, we see that the DCT radiation dose
7:39
index, which I'm going to discuss more detail about the radiation dose index
7:43
later, but the CTDI volume, let's for our case
7:47
right now, look at the relative variation.
7:50
Four milligray, six milligray, and 17 milligray.
7:55
So the question is, do we need this image to diagnose something
7:59
in the patient image?
8:01
If not, we can get away with this mA or this mA.
8:04
That's what the principle of dose standardization or optimization
8:08
comes into picture.
8:10
So regrouping the same principle,
8:14
decreasing the tube current reduces dose linearly.
8:18
Shown here is the tube current setting,
8:21
correspondingly the effective dose setting.
8:25
Again, the effective dose values I'm going to explain in more detail when we
8:29
talk about radiation dose for CT radiation dose, the symmetry, and
8:33
so forth. But for relative comparison, showing here is like
8:36
if this increases linearly,
8:39
50 to 400, eight times it increases
8:43
to 10 millisievert. That's a straightforward linearity.