Interactive Transcript
0:00
The fourth primary factor is pitch.
0:04
The concept of pitch came only after the introduction of
0:08
helical CT scanner or the spiral CT scanner,
0:13
and that is defined as the following.
0:16
This is the table. The patient is moved in the CT table,
0:20
I is called the table feed per rotation, how much
0:24
table is transported through the gantry.
0:28
While this is the X-ray tube, the side view of the X-ray tube and the
0:31
multi-detector CT,
0:33
the pitch is defined as the following.
0:36
Pitch is the ratio of the table feed per
0:39
rotation divided by the X-ray beam width.
0:45
When I say X-ray beam width, if it is a single detector, that is the
0:49
dimension of the detector.
0:50
With the multiple row detector, the width is basically a
0:54
product of the number of detectors multiplied by the
0:58
active channel. That's called the size of each detector,
1:02
multiplying the number of channels we are using, and that gives the beam
1:06
width. Make sense? Now, the radiation dose
1:10
to the patient goes inversely as the pitch value.
1:15
Radiation dose is directly proportional to mAs,
1:18
but is inversely proportional to pitch.
1:22
How do we picture this in mind? Here's an example.
1:26
When pitch is equal to one means the table is
1:30
translated
1:32
in such a speed such that the beam width is
1:36
almost same as the table speed,
1:39
which means it's kind of blocking a Lego box one after the other
1:43
and there's no gap in this one. Imagine a Slinky,
1:47
and if the Slinky is collapsed, brought back to come tightly, there is
1:51
no air gap in between. That is a pitch of one.
1:54
Means the table movement is same as the beam width,
1:59
so every anatomy is covered. That's a pitch of one.
2:03
A pitch greater than one means
2:06
you are moving faster to the table, faster
2:09
than the beam width,
2:12
which means here we have certain gaps developed here,
2:17
and the pitch greater than one means you are extending the
2:20
imaging with reduced patient dose because there are
2:24
certain part of the anatomy is not exposed to radiation dose,
2:28
therefore, you are able to reduce the radiation dose, and
2:32
that's why this pitch value greater than one means the patient
2:36
dose will be less.
2:39
The other aspect in when the Slinky is collapsed
2:43
means if the table is traveling very slowly
2:47
while the beam is larger, then you have lot of overlap in the
2:51
anatomy. That implies a pitch less than one.
2:55
If the pitch is less than one means the dose will be higher,
2:59
and there are scenarios which I'm going to discuss later in cardiac CT,
3:04
where we use a very low pitch value, therefore, the
3:07
patient doses can be higher. That happen because of the
3:11
overlapping and higher patient dose.
3:15
Bringing back to the same simulation here,
3:18
we are now keeping all the techniques same except the
3:22
pitch factor.
3:23
And the pitch factor is changed, its pitch is
3:27
0.64, pitch is 0.83, pitch is
3:30
0.148.
3:32
If we look this one, compared to the CTDI value of 37
3:36
milligray,
3:37
a lower pitch means usually it's a higher value of DC
3:41
and radiation dose.
3:43
That's about 30% higher.
3:46
Whereas a pitch greater than 1.1, which here is
3:49
1.48, the CTDI is now 45%
3:53
lower.
3:55
Now, the question comes how high pitch you can go,
4:00
and there are certain limitation.
4:02
If you open up the Slinky too much, the interpolating
4:06
the data from adjacent area becomes incorrect.
4:10
So it depends on the vendor. Usually, you can go up to
4:14
1.5 pitch value. Beyond that, the error
4:18
creeps in, the axial resolution goes down, therefore, it is not
4:21
advised use.
4:23
So in which clinical scenario you use a pitch greater than one
4:28
or a pitch less than one?
4:31
For typically for most abdominal CT scans,
4:35
where you have the lesser movement of the abdomen and
4:39
everything, we can get away with a pitch greater than one,
4:42
therefore, we can reduce the radiation dose to the patient.
4:46
On the other hand, for cardiac CT, where the
4:50
heart is beating very fast, you tend to do a pitch
4:54
less than one,
4:55
therefore, the retrospective ECG gating techniques tend to
4:59
result in a higher radiation dose to the patient.
5:03
Typically, for a head CT, even though sometime we use a pitch greater
5:07
than one, generally use the pitch of one because of the dense
5:11
skull involved, and most vendor prefer to keep it
5:15
a pitch of equal to one.
5:17
This affects number of things. Pitch will affect the total
5:21
scan time. If it is greater than one, you're going to be covering a larger area.
5:25
It has an impact on the radiation dose.
5:29
It also has an impact on noise resolution.
5:32
Here is a phantom image of showing low contrast resolution.
5:37
If this is a pitch of 0.56, you can see this six millimeters
5:41
object, five millimeter object, and even in fact four millimeter object,
5:45
and if you squint the eyes, you can even see the three millimeter object.
5:49
Now, with the pitch greater than one, which means less
5:53
X-ray photons are contributing the image, you can immediately see the
5:57
object becomes diminishes. The noise dominates,
6:01
therefore, you can hardly see the six millimeter or the five millimeter,
6:05
definitely not the three or smaller than that.
6:08
This is a concept of pitch.
6:10
There is a caveat to the pitch concept.
6:13
Two vendors, Siemens and Philips, uses a concept called
6:17
effective mAs,
6:19
which means they account the pitch inside,
6:23
therefore, if we change the pitch value, the scanner
6:27
will automatically adjust the mA or the scan time
6:31
to get the same effective mAs.
6:34
So
6:36
in this particular vendor's concepts here, the change in the pitch
6:40
does not have any impact on the radiation dose, either increase or
6:44
decrease, because that automatically corrected inside by
6:47
either increasing the mA or decreasing the mA to maintain what is called as
6:52
the effective mAs. This concept is based on
6:55
maintaining image quality independent of pitch setting.