Upcoming Events
Log In
Pricing
Free Trial

Scan Parameters Impacting Radiation Dose and Image Quality - Pitch

HIDE
PrevNext

0:00

The fourth primary factor is pitch.

0:04

The concept of pitch came only after the introduction of

0:08

helical CT scanner or the spiral CT scanner,

0:13

and that is defined as the following.

0:16

This is the table. The patient is moved in the CT table,

0:20

I is called the table feed per rotation, how much

0:24

table is transported through the gantry.

0:28

While this is the X-ray tube, the side view of the X-ray tube and the

0:31

multi-detector CT,

0:33

the pitch is defined as the following.

0:36

Pitch is the ratio of the table feed per

0:39

rotation divided by the X-ray beam width.

0:45

When I say X-ray beam width, if it is a single detector, that is the

0:49

dimension of the detector.

0:50

With the multiple row detector, the width is basically a

0:54

product of the number of detectors multiplied by the

0:58

active channel. That's called the size of each detector,

1:02

multiplying the number of channels we are using, and that gives the beam

1:06

width. Make sense? Now, the radiation dose

1:10

to the patient goes inversely as the pitch value.

1:15

Radiation dose is directly proportional to mAs,

1:18

but is inversely proportional to pitch.

1:22

How do we picture this in mind? Here's an example.

1:26

When pitch is equal to one means the table is

1:30

translated

1:32

in such a speed such that the beam width is

1:36

almost same as the table speed,

1:39

which means it's kind of blocking a Lego box one after the other

1:43

and there's no gap in this one. Imagine a Slinky,

1:47

and if the Slinky is collapsed, brought back to come tightly, there is

1:51

no air gap in between. That is a pitch of one.

1:54

Means the table movement is same as the beam width,

1:59

so every anatomy is covered. That's a pitch of one.

2:03

A pitch greater than one means

2:06

you are moving faster to the table, faster

2:09

than the beam width,

2:12

which means here we have certain gaps developed here,

2:17

and the pitch greater than one means you are extending the

2:20

imaging with reduced patient dose because there are

2:24

certain part of the anatomy is not exposed to radiation dose,

2:28

therefore, you are able to reduce the radiation dose, and

2:32

that's why this pitch value greater than one means the patient

2:36

dose will be less.

2:39

The other aspect in when the Slinky is collapsed

2:43

means if the table is traveling very slowly

2:47

while the beam is larger, then you have lot of overlap in the

2:51

anatomy. That implies a pitch less than one.

2:55

If the pitch is less than one means the dose will be higher,

2:59

and there are scenarios which I'm going to discuss later in cardiac CT,

3:04

where we use a very low pitch value, therefore, the

3:07

patient doses can be higher. That happen because of the

3:11

overlapping and higher patient dose.

3:15

Bringing back to the same simulation here,

3:18

we are now keeping all the techniques same except the

3:22

pitch factor.

3:23

And the pitch factor is changed, its pitch is

3:27

0.64, pitch is 0.83, pitch is

3:30

0.148.

3:32

If we look this one, compared to the CTDI value of 37

3:36

milligray,

3:37

a lower pitch means usually it's a higher value of DC

3:41

and radiation dose.

3:43

That's about 30% higher.

3:46

Whereas a pitch greater than 1.1, which here is

3:49

1.48, the CTDI is now 45%

3:53

lower.

3:55

Now, the question comes how high pitch you can go,

4:00

and there are certain limitation.

4:02

If you open up the Slinky too much, the interpolating

4:06

the data from adjacent area becomes incorrect.

4:10

So it depends on the vendor. Usually, you can go up to

4:14

1.5 pitch value. Beyond that, the error

4:18

creeps in, the axial resolution goes down, therefore, it is not

4:21

advised use.

4:23

So in which clinical scenario you use a pitch greater than one

4:28

or a pitch less than one?

4:31

For typically for most abdominal CT scans,

4:35

where you have the lesser movement of the abdomen and

4:39

everything, we can get away with a pitch greater than one,

4:42

therefore, we can reduce the radiation dose to the patient.

4:46

On the other hand, for cardiac CT, where the

4:50

heart is beating very fast, you tend to do a pitch

4:54

less than one,

4:55

therefore, the retrospective ECG gating techniques tend to

4:59

result in a higher radiation dose to the patient.

5:03

Typically, for a head CT, even though sometime we use a pitch greater

5:07

than one, generally use the pitch of one because of the dense

5:11

skull involved, and most vendor prefer to keep it

5:15

a pitch of equal to one.

5:17

This affects number of things. Pitch will affect the total

5:21

scan time. If it is greater than one, you're going to be covering a larger area.

5:25

It has an impact on the radiation dose.

5:29

It also has an impact on noise resolution.

5:32

Here is a phantom image of showing low contrast resolution.

5:37

If this is a pitch of 0.56, you can see this six millimeters

5:41

object, five millimeter object, and even in fact four millimeter object,

5:45

and if you squint the eyes, you can even see the three millimeter object.

5:49

Now, with the pitch greater than one, which means less

5:53

X-ray photons are contributing the image, you can immediately see the

5:57

object becomes diminishes. The noise dominates,

6:01

therefore, you can hardly see the six millimeter or the five millimeter,

6:05

definitely not the three or smaller than that.

6:08

This is a concept of pitch.

6:10

There is a caveat to the pitch concept.

6:13

Two vendors, Siemens and Philips, uses a concept called

6:17

effective mAs,

6:19

which means they account the pitch inside,

6:23

therefore, if we change the pitch value, the scanner

6:27

will automatically adjust the mA or the scan time

6:31

to get the same effective mAs.

6:34

So

6:36

in this particular vendor's concepts here, the change in the pitch

6:40

does not have any impact on the radiation dose, either increase or

6:44

decrease, because that automatically corrected inside by

6:47

either increasing the mA or decreasing the mA to maintain what is called as

6:52

the effective mAs. This concept is based on

6:55

maintaining image quality independent of pitch setting.

Report

Disclaimer

Faculty

Mahadevappa Mahesh, PhD, FACR, MS, FAAPM, FACMP, FSCCT, FIOMP

Professor of Radiology and Cardiology

Johns Hopkins University School of Medicine

Tags

Physics and Basic Science

Nuclear Medicine

CT