Interactive Transcript
0:00
Scatter radiation is one of the culprit of image quality
0:04
degradation.
0:06
In fact,
0:08
from the time X-rays were developed, X-ray imaging started,
0:12
we have been constantly fighting to develop
0:15
process to eliminate scatter radiation.
0:18
So what is scatter radiation?
0:20
Let's look at the image contrast first.
0:23
Imagine this is an object with two different material, bone and a soft
0:27
tissue. The way we get contrast means the X-rays are
0:31
absorbed differently in these two object.
0:34
Therefore, the signal coming out, if you create an equation like
0:37
this, the signal difference between these two divided
0:41
by the background signal is what we define as a contrast.
0:46
And this contrast is in the absence of scatter.
0:49
There is no scatter, which is purely difference in
0:53
the signal differences between the two.
0:57
Now, the scatter radiation will reduce image contrast,
1:01
and that is influenced by the patient thickness,
1:05
X-ray field area, X-ray energies, and the scatter
1:09
radiation is removed with the Compton scattering, and it is
1:12
removed using grids, which we are going to use it to remove the
1:17
scatter radiation.
1:18
Now, let's say, if there is a scatter,
1:22
the contrast will go
1:24
dramatically with the presence of scatter.
1:27
Before introducing scatter radiation concept, this one, let's look
1:31
at this as a figure.
1:34
In this figure, the X-rays are all primary radiation, which is
1:38
called P, is the primary radiation, passing through the body.
1:42
However, there is one area where the primary radiation
1:45
get deflected, absorbed, and some energy is coming out,
1:50
and this deflected X-rays now reach the same focal
1:54
point.
1:55
Means this is contributed not only by the primary, but also by
1:59
the secondary scatter. And that's what this happen
2:03
is primary... And we use a concept called scatter to
2:07
primary ratio.
2:09
Scatter to primary ratio of three means for every
2:13
primary beam, there is three scatters
2:16
produced, and they are interacting, reaching the same point.
2:20
Imagine now there are 1,000 X-ray photons needed to create
2:24
an image. Now they are all passing through the object, but
2:28
out of this 1,000, almost 300 are now, because they're scattered from different
2:32
area, that's where the fuzziness and the noisiness
2:36
adds, increases. So now how is it happen?
2:40
So the contrast in the absence of any scatter, that is
2:44
the zero scatter to primary ratio,
2:47
let's say is 50%,
2:50
means the contrast between the object, and the background
2:54
is 50%, let's say.
2:56
If there is a scatter,
2:59
for every scatter to primary ratio, if it is one,
3:02
the contrast decreases by 50%,
3:06
almost half.
3:07
If it is scatter primary ratio is two, it further goes down.
3:11
And this is how the contrast changes with respect to the presence
3:15
of scatter.
3:17
An object that has 50% contrast in scatter
3:21
absence has contrast reduced to nearly one-fourth of
3:25
original contrast with the scatter to primary ratio of three.
3:29
In reality, we encounter a scatter to primary ratio of three to four
3:33
easily because of the patient thickness and the field size
3:37
and so forth.
3:39
So now,
3:41
in order to remove the scatter,
3:44
a device was invented, and that's called the anti-scatter
3:47
grid.
3:49
It's basically imagine a mat with a bamboo stick.
3:53
In place between the bamboo stick, you have a lead septum in between,
3:57
and that's how a typical anti-scatter grid looks.
4:01
And the purpose of it is like there is lead septum will
4:04
absorb any scatter coming in different direction.
4:08
Therefore, only primary beam will hit the detector.
4:13
That is idealistic.
4:15
At the same time, when this lead septum will also absorb some
4:19
of the primary beam, therefore, when you have an
4:23
anti-scatter grid introduced, you have to increase the patient
4:27
dose.
4:28
So the grid is placed between the patient and the image receptor,
4:33
and uses geometries, reduce the scatter, and there is
4:37
a concept called as a grid ratio.
4:39
The grid ratio is defined as follows.
4:42
It is ratio, the height of the spectrum, this is the height,
4:46
means the length of the septum, and divided by the width.
4:50
The spacing between the septum is called the grid ratio.
4:55
That can typically go from 10 to one or 12 is to one,
4:59
five is to one, and they're usually about 60 line
5:03
per centimeter. That's why we don't see the grid as an empty space.
5:07
It's all compacted in the carbon fiber.
5:11
Now,
5:12
here are the typical grid ratio.
5:16
For radiography and fluoroscopy, especially for table top
5:19
radiography, the grid ratio is eight is to one,
5:24
which means the height of the septum is eight, while the
5:27
width is one,
5:29
whatever unit you use.
5:31
And if you can notice here, they are aligned in such a way, they are
5:35
all focused towards the X-ray focus spot to
5:39
minimize absorption of the primary beam.
5:43
So radiography is eight is to one,
5:46
and then it goes up to or 10 is to one or 12 is to
5:50
one.
5:51
The larger grid ratio is typically used in chest
5:54
radiography because chest radiographs are taken at a
5:58
longer distance compared to the table top radiograph.
6:02
The tabletop radio X-rays are done at 40 inches,
6:06
whereas chest X-rays are done at 72 inches, therefore,
6:10
you can use 12 is to 1 grid.
6:13
In mammography, we use a smaller grid, five is to one or four
6:17
is to one, because the distance between the source and the
6:21
detector is much lesser than radiography or fluoroscopy.
6:25
Because of the introduction of this grid, we have a new concept
6:29
called Bucky.
6:31
Bucky factor is basically the ratio of the
6:35
entrance exposure with grid to the entrance exposure
6:39
without grid.
6:41
So when there's no grid, the Bucky factor is one.
6:45
But if there is a grid, then the dose with the grid has to be higher,
6:50
therefore the Bucky factor will be higher.
6:52
It can range anywhere from two to four, means when you're using
6:56
grid, the dose required to create an image is four times
6:59
higher than without grid. So there is a trade-off.
7:03
Do you want a grid and increase the dose to remove the scatter,
7:08
or you want to remove the grid, reduce the dose, but you are
7:12
encountered with the noise and scatter radiation?
7:15
There are certain clinical scenario where there is no need for
7:19
grid,
7:20
and that's in pediatric case, where the object is much smaller,
7:25
therefore the scatter is lesser. Therefore, in order to keep the radiation
7:28
dose smaller, we remove the grid in pediatric cases.
7:34
Here, an example shown, the scatter to primary beam
7:37
varies with respect to patient thickness.
7:40
If it is a 30-centimeter patient thickness, abdominal thickness, look at
7:44
here, for a 10-centimeter field of view, the
7:48
scatter radiation is about more than five.
7:51
If it is 20 centimeter or 30 centimeter, which is a normal average
7:55
abdominal size, that scatter primary ratio is about six,
7:59
not three. That's what it determines the Bucky factor.
8:04
There is another way to remove the scatter also,
8:07
by simply moving the patient or the detector
8:11
further away from the object.
8:14
This way, the air gap will remove some of the scatter going
8:18
away. They don't hit the detector,
8:21
and this is called air gap methodology of scatter
8:24
radiation. But in this case, the object get
8:28
magnified, so we are not desired in X-rays, so we are trying
8:32
to keep the grid to remove the scatter radiation.
8:35
But one area where we like to use these opportunities,
8:39
in mammography, when they do a mammogram image and you
8:43
need to magnify the image, the breast is placed at a higher
8:47
distance on a breast plate, magnification plate,
8:51
allowing air to use as a way to reduce
8:55
scatter radiation, there is no grid, and also to minimize the radiation
8:59
dose to the patient. So one area which air gap
9:02
methodology is applied is in mammography.