Interactive Transcript
0:01
Let's look at the impact of X-ray beam hardening.
0:04
What is beam hardening?
0:06
When X-rays are produced from an X-ray tube, the X-rays are
0:10
all different energies. There are some low-energy X-rays, some very high-energy
0:14
X-rays. The maximum energy produced is defined by the tube
0:18
voltage we set. But a lot of the low energy is useless for
0:22
image creation. It basically increases the radiation exposure
0:26
to patient.
0:27
So we need to remove some of them,
0:29
and that's done, we are process called as
0:33
hardening the X-ray beam, called as beam hardening.
0:36
And it is done as follows.
0:39
So, the soft X-rays get absorbed on the skin
0:42
surface and do not contribute to formation.
0:47
Therefore, you want to harden the beam to remove the
0:50
soft X-rays and harden the beam so that it has
0:54
sufficient energy to penetrate patient.
0:57
So,
0:59
here is the photon energy and HVL increases.
1:02
So here is the first interaction, all energy range shown by
1:06
different frequencies and wavelength here.
1:09
Lot of the low wavelength is absorbed.
1:13
As you go more and more thickness is placed in between, you
1:17
get more of an average higher energy.
1:20
And that can be shown in this particular spectrum.
1:24
So here is again, this is a typical X-ray spectrum,
1:28
and if I look at here, I can say the tube voltage is 100
1:31
kVp because the maximum energy produced is
1:35
100 kV, we are not beyond that. In this
1:39
particular target or material, there was a characteristic around this
1:43
area, around 60 to 72 here. It may be due to the X-ray.
1:47
So now you calculate what is called as an average energy.
1:51
This is called the peak energy, and if you take all the X-rays
1:55
of different photon,
1:57
number of X-rays in different energy, you average it, we get what is called as
2:01
average energy.
2:03
The average energy is usually between one third to
2:07
two third of the maximum energy. So if it is 100
2:11
kVp, I can confidently say the average energy of
2:15
the X-ray coming out is around 33 to
2:19
66 keV.
2:21
What I'm trying to show here is like the spectrum already has blocked off
2:25
a lot of this low energy, and that's done by the window of the
2:28
X-ray tube or the glass casing, which absorb a lot of the low energy.
2:33
Now, for between these two, it's done at 100 kV,
2:37
250 mA, everything same, except now you add
2:41
one millimeter of aluminum at the point of X-rays coming out of
2:45
the X-ray tube.
2:47
Immediately what happen is like the average energy
2:51
will increase.
2:53
Means the energy coming out is now bit more stronger,
2:58
more penetrable, and it can penetrate more larger body,
3:02
and that's where the shift happened to the right side.
3:05
But notice also there is a intensity goes down because
3:09
some of the photons are lost by absorption. That's why we get this type.
3:14
Now, shown here is no filtration
3:18
with one millimeter filtration, and now with a two and a half
3:22
millimeter of aluminum added, it can go further down,
3:26
and this is with the copper is even more further.
3:29
Intensity goes down, but it's the shift towards the right side.
3:34
Then one can say you can keep on adding more and more material.
3:37
If we add more and more filter material, if the average energy
3:41
is very high, that will begin to impact the image
3:45
contrast.
3:47
Bremsstrahlung process produces wide spectrum of energy, which is that's why
3:51
we call it polyenergetic X-ray beam.
3:54
The beam hardening will shift off X-ray spectrum to
3:58
higher effective energies due to removal of lower
4:02
energy X-ray photon.
4:05
In one slide, basically, the average energy
4:09
keeps decreasing or increasing to the higher level, but the
4:12
intensity will decrease. But this is the
4:16
impact of the beam hardening
4:19
as shown here.
4:21
How does it matter?
4:23
That will directly impact on the image contrast.
4:27
On the upper side, here shown is an average size patient.
4:31
The thickness is about 23 centimeter.
4:33
Anywhere from 20 to 30 centimeter is considered as average.
4:37
The distance between the source and the receptor is 100 centimeter.
4:41
Now, in this scenario, for a 23 centimeter,
4:44
if you set the tube voltage at 50 kV, this
4:48
exposure is 1,656 mR.
4:52
That is the amount of radiation you want to produce or amount of X-rays
4:56
needed to create an image.
4:59
Lower kV means higher dose to the patient.
5:02
It can produce higher contrast, but it will also produce higher patient
5:06
dose.
5:07
Let's say if you go down the list, at 140 kV, the
5:11
entrance X-rays is only 153 mR,
5:15
but that will lose the contrast, showing the contrast differentiating with the
5:19
tissue.
5:20
Therefore, for diagnostic radiology, there's a very limited
5:24
range of kV we can use and still
5:28
producing a good image quality.
5:30
So here is the diagram show the dose
5:33
or contrast percentage, which is the kVp.
5:37
If you look at this, this is the entrance dose,
5:40
and this is the contrast agent, and this is where the dose to the
5:44
patient goes when you're imaging a one-millimeter bone chip.
5:49
Back again to demonstrate, we are working in the range of
5:53
10 to 100 kV for diagnostic imaging.
5:56
The common point is Compton scattering and the photoelectric
6:00
absorption.
6:02
The HVL is defined as follows,
6:05
is given as in a narrow beam geometry, if X-ray is passing
6:09
through, you keep putting the material until the intensity
6:13
goes by 50%, and that material is called half value
6:17
layer.
6:18
Why is that important? I'm trying to term in terms of the patient
6:22
thickness.
6:23
Let's say for a patient thickness and a soft tissue of
6:27
37 millimeter,
6:29
just about four centimeter,
6:32
for a 70 kV X-rays,
6:35
it only requires 3.37 millimeter to cut the intensity
6:39
by half.
6:41
For a 140 kV gammas, it requires about 44 millimeter.
6:45
What I'm trying to show you is
6:47
in a tissue,
6:49
X-rays are absorbed.
6:51
Half of the X-rays incident on the tissues are
6:54
absorbed every four centimeter in distance.
6:59
So approximately four centimeter means if a patient abdominal
7:03
thickness is 20 centimeter,
7:05
after every 40 millimeter, four centimeter, half of the
7:10
X-rays are remaining. So by the time X-rays comes out, it's
7:13
approximately 15% of the incident radiation,
7:17
which is used for image creation. So that's where the challenge is.
7:21
One other point which I want to bring here is
7:24
for lead. You can see here the lead,
7:27
every 0.2 millimeter will absorb by half of the radiation.
7:31
That's one of the reason why we use lead apron, because any
7:35
scattering come will be completely absorbed by the lead.
7:40
How does the attenuation makes the image contrast?
7:44
The image contrast is due to variation in attenuation.
7:48
In this case, the thickness is same
7:52
in both the situation,
7:54
but this is a wax and this is aluminum.
7:57
The lower attenuation material, wax, has lower contrast,
8:01
whereas this has a higher contrast because lot of it is absorbed
8:05
compared to the background, therefore, this is much more differentiated.