Upcoming Events
Log In
Pricing
Free Trial

Impact of X-ray Beam Hardening

HIDE
PrevNext

0:01

Let's look at the impact of X-ray beam hardening.

0:04

What is beam hardening?

0:06

When X-rays are produced from an X-ray tube, the X-rays are

0:10

all different energies. There are some low-energy X-rays, some very high-energy

0:14

X-rays. The maximum energy produced is defined by the tube

0:18

voltage we set. But a lot of the low energy is useless for

0:22

image creation. It basically increases the radiation exposure

0:26

to patient.

0:27

So we need to remove some of them,

0:29

and that's done, we are process called as

0:33

hardening the X-ray beam, called as beam hardening.

0:36

And it is done as follows.

0:39

So, the soft X-rays get absorbed on the skin

0:42

surface and do not contribute to formation.

0:47

Therefore, you want to harden the beam to remove the

0:50

soft X-rays and harden the beam so that it has

0:54

sufficient energy to penetrate patient.

0:57

So,

0:59

here is the photon energy and HVL increases.

1:02

So here is the first interaction, all energy range shown by

1:06

different frequencies and wavelength here.

1:09

Lot of the low wavelength is absorbed.

1:13

As you go more and more thickness is placed in between, you

1:17

get more of an average higher energy.

1:20

And that can be shown in this particular spectrum.

1:24

So here is again, this is a typical X-ray spectrum,

1:28

and if I look at here, I can say the tube voltage is 100

1:31

kVp because the maximum energy produced is

1:35

100 kV, we are not beyond that. In this

1:39

particular target or material, there was a characteristic around this

1:43

area, around 60 to 72 here. It may be due to the X-ray.

1:47

So now you calculate what is called as an average energy.

1:51

This is called the peak energy, and if you take all the X-rays

1:55

of different photon,

1:57

number of X-rays in different energy, you average it, we get what is called as

2:01

average energy.

2:03

The average energy is usually between one third to

2:07

two third of the maximum energy. So if it is 100

2:11

kVp, I can confidently say the average energy of

2:15

the X-ray coming out is around 33 to

2:19

66 keV.

2:21

What I'm trying to show here is like the spectrum already has blocked off

2:25

a lot of this low energy, and that's done by the window of the

2:28

X-ray tube or the glass casing, which absorb a lot of the low energy.

2:33

Now, for between these two, it's done at 100 kV,

2:37

250 mA, everything same, except now you add

2:41

one millimeter of aluminum at the point of X-rays coming out of

2:45

the X-ray tube.

2:47

Immediately what happen is like the average energy

2:51

will increase.

2:53

Means the energy coming out is now bit more stronger,

2:58

more penetrable, and it can penetrate more larger body,

3:02

and that's where the shift happened to the right side.

3:05

But notice also there is a intensity goes down because

3:09

some of the photons are lost by absorption. That's why we get this type.

3:14

Now, shown here is no filtration

3:18

with one millimeter filtration, and now with a two and a half

3:22

millimeter of aluminum added, it can go further down,

3:26

and this is with the copper is even more further.

3:29

Intensity goes down, but it's the shift towards the right side.

3:34

Then one can say you can keep on adding more and more material.

3:37

If we add more and more filter material, if the average energy

3:41

is very high, that will begin to impact the image

3:45

contrast.

3:47

Bremsstrahlung process produces wide spectrum of energy, which is that's why

3:51

we call it polyenergetic X-ray beam.

3:54

The beam hardening will shift off X-ray spectrum to

3:58

higher effective energies due to removal of lower

4:02

energy X-ray photon.

4:05

In one slide, basically, the average energy

4:09

keeps decreasing or increasing to the higher level, but the

4:12

intensity will decrease. But this is the

4:16

impact of the beam hardening

4:19

as shown here.

4:21

How does it matter?

4:23

That will directly impact on the image contrast.

4:27

On the upper side, here shown is an average size patient.

4:31

The thickness is about 23 centimeter.

4:33

Anywhere from 20 to 30 centimeter is considered as average.

4:37

The distance between the source and the receptor is 100 centimeter.

4:41

Now, in this scenario, for a 23 centimeter,

4:44

if you set the tube voltage at 50 kV, this

4:48

exposure is 1,656 mR.

4:52

That is the amount of radiation you want to produce or amount of X-rays

4:56

needed to create an image.

4:59

Lower kV means higher dose to the patient.

5:02

It can produce higher contrast, but it will also produce higher patient

5:06

dose.

5:07

Let's say if you go down the list, at 140 kV, the

5:11

entrance X-rays is only 153 mR,

5:15

but that will lose the contrast, showing the contrast differentiating with the

5:19

tissue.

5:20

Therefore, for diagnostic radiology, there's a very limited

5:24

range of kV we can use and still

5:28

producing a good image quality.

5:30

So here is the diagram show the dose

5:33

or contrast percentage, which is the kVp.

5:37

If you look at this, this is the entrance dose,

5:40

and this is the contrast agent, and this is where the dose to the

5:44

patient goes when you're imaging a one-millimeter bone chip.

5:49

Back again to demonstrate, we are working in the range of

5:53

10 to 100 kV for diagnostic imaging.

5:56

The common point is Compton scattering and the photoelectric

6:00

absorption.

6:02

The HVL is defined as follows,

6:05

is given as in a narrow beam geometry, if X-ray is passing

6:09

through, you keep putting the material until the intensity

6:13

goes by 50%, and that material is called half value

6:17

layer.

6:18

Why is that important? I'm trying to term in terms of the patient

6:22

thickness.

6:23

Let's say for a patient thickness and a soft tissue of

6:27

37 millimeter,

6:29

just about four centimeter,

6:32

for a 70 kV X-rays,

6:35

it only requires 3.37 millimeter to cut the intensity

6:39

by half.

6:41

For a 140 kV gammas, it requires about 44 millimeter.

6:45

What I'm trying to show you is

6:47

in a tissue,

6:49

X-rays are absorbed.

6:51

Half of the X-rays incident on the tissues are

6:54

absorbed every four centimeter in distance.

6:59

So approximately four centimeter means if a patient abdominal

7:03

thickness is 20 centimeter,

7:05

after every 40 millimeter, four centimeter, half of the

7:10

X-rays are remaining. So by the time X-rays comes out, it's

7:13

approximately 15% of the incident radiation,

7:17

which is used for image creation. So that's where the challenge is.

7:21

One other point which I want to bring here is

7:24

for lead. You can see here the lead,

7:27

every 0.2 millimeter will absorb by half of the radiation.

7:31

That's one of the reason why we use lead apron, because any

7:35

scattering come will be completely absorbed by the lead.

7:40

How does the attenuation makes the image contrast?

7:44

The image contrast is due to variation in attenuation.

7:48

In this case, the thickness is same

7:52

in both the situation,

7:54

but this is a wax and this is aluminum.

7:57

The lower attenuation material, wax, has lower contrast,

8:01

whereas this has a higher contrast because lot of it is absorbed

8:05

compared to the background, therefore, this is much more differentiated.

Report

Faculty

Mahadevappa Mahesh, PhD, FACR, MS, FAAPM, FACMP, FSCCT, FIOMP

Professor of Radiology and Cardiology

Johns Hopkins University School of Medicine

Tags

X-Ray (Plain Films)

Physics and Basic Science