Interactive Transcript
0:00
So in order to create an X-ray signal, we need an X-ray tube.
0:04
The fundamental principle of an X-ray tube are the following.
0:08
These are the requirement. The X-ray tube need to have an evacuated chamber.
0:14
It needs to have a source of free electron,
0:17
because those are the electron which is going to interact with the target surface,
0:22
which we call it as anode. And that target surface has to be a heavy
0:26
metal because it should not be easily melted out or
0:30
vaporized. So that again defines the choice of the target
0:34
material we choose. We also need to have a way of accelerating the
0:39
electron from the cathode to anode and hitting it.
0:44
And in the process, we also have a means to getting rid of the heat,
0:49
because the X-ray tube, the process is quite
0:53
inefficient because 99% of the time it's the heat which is
0:57
produced, only 1% of it is X-rays produced.
1:01
So there's a challenge to push out or remove the heat as soon as possible.
1:07
This is one of the fundamental X-ray tube design.
1:11
Pretty much from old days to nowadays, it's still the principal
1:15
schematic diagram works. So what are the different factors in this particular
1:20
X-ray tube? The X-ray tube is a filament here.
1:24
This is made up of a tungsten filament, and it is heated by a power
1:29
supply. It's almost like a light bulb.
1:32
When the light is turned on, the current passes through the filament and the
1:35
filament will glow, and that's exactly here.
1:40
The filament will release electron.
1:42
We call this photoelectron. Electrons are negatively
1:46
charged. They are now attracted towards the positively charged
1:51
anode, and the way it is done is maintained by a voltage.
1:56
That's the one which determines the tube voltage we set on X-rays.
2:01
And that amount of current, the electron produced, is determined by the tube
2:04
current, and that is called the mA, which is the protocol setting of the mA of the
2:09
protocols. So these electron, they are
2:14
accelerated towards the anode, and when they hit that anode,
2:19
that's when X-rays are produced, both Bremsstrahlung radiation or characteristic
2:24
X-rays. And some of it is absorbed here, some of it is coming out
2:29
of a window here, which is what is positioned on the patient
2:33
to capture this X-ray passing through the patient to get an image.
2:38
Apart from this, there is also what is called as the anode.
2:41
On top of it is a tungsten target is embedded in the anode, and this
2:46
rotates or different methods are applied to take the heat off.
2:51
And this is the fundamental diagram of an X-ray tube.
2:56
Here are some images of X-ray tube.
2:59
This is the early cold cathode X-ray tube,
3:03
probably back in 1920s. This is one of those 1940s X-ray tube housed
3:07
in a glass casing, which is no longer the case because the X-ray tube are
3:12
well kept inside a sealed and a lead capsulated chamber
3:17
so that you're trying to avoid any of these radiations coming out of the X-ray
3:21
tube to be captured, absorbed. We call this a leakage radiation.
3:27
Now, if you slice through the X-ray tube, which is currently well housed in a
3:32
very big container, this is a side view of an X-ray tube.
3:38
So you have a cathode, which is the filament element here, and you have
3:42
an anode, which is here, which is actually fixed on a
3:47
rotating cup, and this rotates quite fast.
3:50
By doing so, you can take the heat out of the X-ray production.
3:55
A cathode, anode, you have a rotor, which rotates at almost 10,000
4:00
revolution per minute. You have a glass and a metal envelope usually
4:05
covering everything inside, and then you have a tube housing.
4:09
The whole thing is housed inside this.
4:12
And this housing, this is a X-ray tube, and this is a tube housing, can
4:16
have air passing through to take the heat off, or it can be immersed
4:21
in oil, because oil is also another way of conducting the
4:26
heat out and so forth.
4:28
And this is the output port where you are directing the X-rays to come out.
4:35
If you look examine it closely with respect to cathode, shown here is the
4:39
filament. This is the filament. This is called a filament cup or a
4:44
focusing cup. There are two different sizes of a filament, and I'll
4:49
tell you why, because we have different filaments, and that will also
4:53
determine what the best spatial resolution you can get in imaging.
4:58
Smaller the filament, greater the spatial resolution.
5:01
However, smaller filament cannot operate continuously, so for
5:06
a fluoroscopy or interventional radiology, we use a larger
5:10
filament. Apart from that, you have filament made up of tungsten
5:14
wire, because tungsten wire has a very high melting point, so it can
5:19
sustain large amount of heat it produced.
5:23
The filament current is typically three to six mA, and that when it does,
5:28
you have a process called thermionic emission, which releases the electron out of
5:33
the filament.
5:34
You have a focusing cup. Focusing cup is basically to focus this
5:39
electron towards the anode. And most X-ray tube have two X-ray filament,
5:44
and there may be X-ray tube which has three filaments in some special procedure.
5:49
So if it is not focused, these electron can wander around not
5:53
going close towards the anode. If it is focused, it will focus in a
5:58
beam of electron towards the anode, which will result in a higher spatial
6:02
resolution, and most of it is efficiently utilized.
6:07
So there is also a structure on the anode. The anode looks like this.
6:12
It is called a target electrode, which is maintained at a positive
6:17
potential with respect to cathode.
6:20
Electron from cathode, upon impact, transfer energy by
6:24
collision and radiative interaction with anode producing
6:29
heat and X-rays. So when these electron hit
6:33
the anode here at a certain speed,
6:37
99% of the time there's a heat produced, and 1% of the time there's an X-ray
6:42
produced, and this X-ray is then captured towards the outside of the
6:46
window, which is what is used for imaging.
6:50
The most common target is tungsten because it has very high melting
6:55
point and high atomic number. And this anode
7:00
is fixed into a rotor, so the rotor can rotate faster so that the
7:04
electron don't hit only one spot on the anode.
7:08
If that happens, that anode can get burnt off or get a pit, which can cause more
7:13
artifact. So when it is rotating, the X-rays are exposing a track
7:18
on the particular anode, and that is created.
7:22
And we're going to discuss a little bit more about the focal spot and so forth
7:25
later.
7:27
One other key factor in radiography is called the heel effect.
7:33
Heel effect is the heel similar to our heel in the foot.
7:37
Imagine the X-rays are produced here.
7:40
The electrons are hitting the anode, and the electron creating X-rays, X-rays
7:45
coming out in all different way. Some of it is also
7:49
absorbed by the anode itself.
7:53
Because of that, the X-ray intensity measured in this
7:57
direction is much lesser than the X-ray measured in this
8:01
direction. And we call this as a heel effect.
8:05
Means the heel of the foot is absorbing some of the X-rays, therefore, the
8:10
intensity in that direction is lesser compared to the other direction.
8:14
It's the maximum here, lesser here, and it's less on the other side.
8:20
So X-ray emitted on the anode side
8:24
must pass through a greater thickness of anode, which results in a
8:29
reduced intensity.
8:31
Heel effect is less important with larger focal spot size.
8:35
If the focal spot is very small, you can see this heel effect.
8:38
So you can see here, shorter distance to exit the beam, there is less
8:43
attenuation it comes out, whereas here, the X-ray produced, they're also
8:48
absorbed by the anode material itself, so therefore, the intensity is less.
8:53
Why I'm sharing this heel effect? This becomes really important when we are
8:57
discussing mammography X-ray tube and the way it's positioned with respect to the
9:02
patient. For most radiography, the heel effect is now negated to some
9:07
extent in terms of the averaging the signals in a digital detector.